π Process Input β Sample Composition
SourcePipeline / Process Stream
Carrier GasHelium (He)
Inlet Pressure2.8 bar
Component Composition (%)
CHβ75%
CβHβ12%
CβHβ6%
COβ4%
Nβ3%
Total100%
π¨ Performance & Alarms
β
Separation Quality: Good
β
Column Temperature: Normal
β
Carrier Gas Flow: Normal
β
Sample Load: Normal
| Oven Status | HEATING |
| Column Pressure | 2.8 bar |
| Detector | TCD β READY |
| Run Status | STANDBY |
| Inject Count | 0 |
| Run Time | β |
βοΈ GC System β Process Flow Diagram
π Real-Time Chromatogram OutputRun Time: 0.0 min
π§ Learning & Reference Panel
GC WORKING PRINCIPLE β STEP BY STEP
1
Sample Introduction: A precise volume of gas sample (typically 0.25β1 mL) is introduced via a 6-port switching valve. The valve switches from LOAD to INJECT position, pushing the sample slug into the carrier gas stream.
2
Carrier Gas Transport: An inert gas (Helium or Nitrogen) acts as the mobile phase. It carries sample molecules from the injection point through the entire column at controlled pressure and flow rate.
3
Column Separation: Inside the heated oven, molecules interact with the stationary phase coating. Lighter, less polar molecules (CHβ, Nβ) interact weakly and exit first. Heavier molecules (CβHβ) interact more strongly and exit later.
4
Thermal Effect: Column temperature controls interaction strength. Higher temp = shorter retention times but reduced resolution. Temperature programming can optimize both speed and separation quality.
5
Detection: The TCD (Thermal Conductivity Detector) measures changes in electrical resistance of a heated filament as each component passes through. FID is used for hydrocarbons. Each component produces a unique electrical signal peak.
6
Chromatogram Output: The data system plots signal intensity vs. time. Each peak represents one component. Peak retention time = component identity. Peak area = concentration (after calibration).
7
Quantification: Peak areas are compared to known calibration standards. Results are expressed as mol%, vol%, or mass%. Calibration is performed with certified reference gas mixtures (CRGM).
Carrier GasStationary PhaseMobile PhaseRetention TimeResolutionPeak AreaSelectivityEfficiencyvan Deemter
Retention Time (tR): Time from injection to peak maximum. Unique fingerprint for each compound under fixed conditions. Used for qualitative identification.
Resolution (R): Degree of peak separation. R = (tR2βtR1)/(0.5Γ(w1+w2)). R > 1.5 = baseline separation. R < 1.0 = peaks overlap.
Stationary Phase: Polydimethylsiloxane (DB-1, HP-1) coating on column wall. Interacts with sample via van der Waals and dipole forces. Phase polarity determines selectivity.
Peak Area: Proportional to component concentration. Must be calibrated with CRGM. Area% method common for natural gas analysis.
Column Efficiency (N): Number of theoretical plates. N = 5.54Γ(tR/wΒ½)Β². Higher N = narrower peaks = better separation. Typical capillary: 20,000β100,000 plates.
van Deemter Equation: H = A + B/u + Cu. Describes height equivalent of theoretical plate vs. carrier gas velocity. Optimal flow rate minimizes H and maximizes N.
Partition Coefficient (K): K = Cs/Cm (concentration in stationary/mobile phase). High K = long retention. Determines relative order of elution for each component.
βοΈ INSTALLATION & COMMISSIONING
- Verify all connections for leak tightness using appropriate leak detection fluid or He detector
- Condition new column at 10β15Β°C below max temp for β₯30 min before first use
- Flush system with carrier gas for minimum 30 minutes before introducing sample
- Calibrate with certified reference gas mixture (CRGM) traceable to NPL/NIST
- Verify detector response linearity across full composition range expected
π OPERATING BEST PRACTICES
- Maintain carrier gas purity β₯99.999% (5.0 grade) for TCD applications
- Keep column head pressure stable β fluctuations cause retention time drift
- Perform daily calibration check (span and zero) per ISA-37.1 requirements
- Use isothermal temperature for consistent retention time reproducibility
- Avoid sample overloading β inject minimum volume required for detection
- Monitor baseline noise and drift β abnormal drift indicates column contamination
π§ MAINTENANCE
- Replace sample valve seals/rotor every 6β12 months depending on cycle count
- Clean or replace sample loop if contamination or carryover is detected
- Check carrier gas cylinder pressure daily β maintain adequate supply
- Inspect and clean detector filaments per manufacturer schedule
- Record all maintenance activities in GC logbook for audit traceability
- Perform column aging (baking) periodically to remove high-boiling contaminants
β οΈ TROUBLESHOOTING
- No peaks: Check carrier gas supply, valve position, detector power
- Peak overlap: Reduce column temp, reduce flow rate, or use longer column
- Retention time drift: Check column pressure, temperature stability, carrier gas purity
- Tailing peaks: Active sites on column β bake out column or replace
- Negative peaks (TCD): Component thermal conductivity similar to carrier gas
- Split peaks: Two-phase injection, condensation in sample system
π STANDARDS & CODES
- ASTM D1945 β Natural gas analysis by GC
- ASTM D1946 β Refinery gas analysis
- ISO 6974 β Natural gas composition determination
- GPA 2261 β LNG/NGL composition analysis
- ISA-37.1 β Instrument performance standard
π― QUALITY ASSURANCE
- Replicate injections (nβ₯3) required for statistical confidence in results
- Acceptable RSD (relative standard deviation) β€0.5% for custody transfer GC
- Use two-point calibration (zero + span) for linear detector response verification
- Archive all calibration records and chromatograms per site data retention policy
- Participate in inter-laboratory comparison (round-robin) programs annually
- Validate any method changes against certified reference material
β½ OIL & GAS
βΆ
Natural Gas Metering: Calorific value, Wobbe Index, and compressibility factor calculation for custody transfer per ISO 6976. GC feeds real-time composition data to flow computers.
βΆ
Offshore Platform: Wellhead gas composition for reservoir management, flare gas emissions reporting, produced water hydrocarbon content monitoring.
βΆ
Pipeline: Gas quality monitoring for tariff compliance, odorization verification (THT/mercaptans), interoperability checks at interconnection points.
π§ LNG & REFINERY
βΆ
LNG Terminal: Methane number calculation for marine fuel, boil-off gas composition monitoring, regasification product quality verification per GIIGNL standards.
βΆ
Refinery: Reformate composition (BTX), FCC off-gas analysis, C3/C4 splitter control, product specification verification, benzene/toluene/xylene purity.
βΆ
Crude Distillation: Naphtha true boiling point (TBP) analysis, light-ends recovery optimization, crude oil characterization by simulated distillation (SimDist).
π PETROCHEMICAL
βΆ
Polymer Feed: Ethylene purity β₯99.9% for polymerization feed, propylene purity verification, vinyl chloride monomer (VCM) analysis, trace impurities.
βΆ
BTX Plants: Benzene, toluene, xylene purity monitoring for aromatic extraction units. Sulfolane solvent quality tracking.
βΆ
Emissions Monitoring: VOC analysis, greenhouse gas composition (COβ, CHβ, NβO), continuous emissions monitoring system (CEMS) per EPA Method 18.